Coriolis Flow Meter Calibration

Coriolis flow meter calibration is performed to ensure the measurement accuracy of mass flow meters. Mass flow meters are high-precision measuring instruments, and maintaining high measurement accuracy is critical in applications such as industrial processes and trade settlement. Coriolis flow meter calibration involves comparing the meter’s readings with those of a standard meter to determine and correct the mass flow meter’s accuracy.
Sino-Inst’s Coriolis flow meters are pre-calibrated with water before leaving the factory, and the accuracy of the flow coefficient calibration is specified in the factory report. After installation, users typically need to perform only one task: zero adjustment. If a weight comparison confirms that the readings are consistently too high or too low, the flow coefficient can be adjusted.
Factory Calibration
Our factory holds a Chinese Measurement Type Approval Certificate, and our mass flow meters are CE and ATEX certified. Every mass flow meter we customize for our customers is calibrated with water before shipment.
The mass flow meter is mounted on our calibration bench, and water is passed through it at several different flow rates to compare the meter’s readings with a reference standard. The result of this comparison is the flow coefficient.
The mass flow meter’s transmitter uses the flow coefficient to convert the sensor’s raw signal into kg/h. The coefficient varies for each flow meter because no two measuring tubes have exactly the same stiffness.
Our factory then issues a calibration report detailing information such as the accuracy, zero point, flow coefficient, and temperature coefficient for each calibration point.
Zero Adjustment
In addition to factory calibration, which ensures the accuracy of the mass flow meter, users can also maintain measurement accuracy through zero adjustment during operation. Zero adjustment (also called zero calibration) refers to the process of adjusting the instrument so that its output signal (e.g., 4 mA) corresponds to zero when the actual flow rate of the mass flow meter is zero.
When is zero adjustment necessary? We recommend:
- When the meter is newly installed and used for the first time;
- When the meter has been removed and reinstalled, or its position has been changed;
- When the medium has been changed;
- When the operating temperature differs significantly from that at the time of the last zero adjustment;
- When the valve is fully closed but the meter still displays flow;
- When readings are noticeably incorrect at low flow rates;
There are two situations under which zero adjustment must never be performed:
- While the medium is still flowing. The meter will interpret the flow at this point as zero, causing all subsequent readings to be too low.
- When the pipe is not fully filled with liquid or contains air bubbles. In this case, the measured zero point is inaccurate.
Every mass flow meter, regardless of size, has a “zero-point stability” specification, which indicates the maximum allowable fluctuation of the zero point. This fluctuation is a fixed value; the smaller the flow rate, the larger the proportion it represents.
| Size | Max. flow (kg/min) | Zero-point stability (kg/min) | Zero error exceeds 0.1% below |
|---|---|---|---|
| DN1 (1/12″) | 0.2 | 0.0002 | 0.2 kg/min |
| DN2 (1/10″) | 1.6 | 0.0005 | 0.5 kg/min |
| DN3 (1/8″) | 3 | 0.001 | 1 kg/min |
| DN4 (1/6″) | 5 | 0.001 | 1 kg/min |
| DN8 (1/4″) | 20 | 0.002 | 2 kg/min |
| DN15 (1/2″) | 60 | 0.006 | 6 kg/min |
| DN20 (3/4″) | 100 | 0.010 | 10 kg/min |
| DN25 (1″) | 200 | 0.010 | 10 kg/min |
| DN40 (1-1/2″) | 450 | 0.020 | 20 kg/min |
| DN50 (2″) | 650 | 0.025 | 25 kg/min |
| DN80 (3″) | 2000 | 0.050 | 50 kg/min |
| DN100 (4″) | 3000 | 0.100 | 100 kg/min |
| DN150 (6″) | 12000 | 0.200 | 200 kg/min |
| DN200 (8″) | 18000 | 0.300 | 300 kg/min |
| DN250 (10″) | 29000 | 0.750 | 750 kg/min |
Take a 1-inch meter as an example. At 100 kg/min, the zero point adds only 0.01% of error. At 2 kg/min, it adds 0.5%, which is more than the accuracy class itself. If your flow rate is often below the value in the last column, the zero point must be adjusted properly.
Preparations before zero adjustment:
- Power on the flowmeter to warm it up, and let the medium flow for a while. Allow the sensor temperature to approach its normal operating temperature.
- Ensure the pipe is fully filled with medium and free of air bubbles.
- Close the valves—start with the downstream valve, then the upstream valve. Closing the downstream valve first helps maintain full pipe pressure. Both valves must be tightly closed; if either leaks, the medium will continue to flow.
- Wait 3 to 5 minutes.
- The pipeline must be free of vibration. It is best to stop any nearby pumps or agitators first.
- Check whether the DP value is stable. On the main screen, press the ▲ key to switch to the DP display page and monitor it closely for a moment. If the value fluctuates by no more than 10, it indicates that the pipeline has settled and you can proceed with zero adjustment. If the fluctuation exceeds this limit, go back and check the previous steps.
Zero Adjustment Procedure
There are three buttons on the meter: SET on the left, ▶ in the middle, and ▲ on the right.
- Press and hold SET to enter the user menu.
- Enter the password. The password is listed in the manual that came with the meter; if you cannot find it, please contact us.
- Press SET to scroll through the pages one by one until the “Zero Calibration” page appears.
- Press ▶ to move the cursor to Y, then press SET to confirm.
- The screen will ask again; move the cursor to Yes and press SET.
- Wait approximately 30 seconds. Do not touch the meter or operate any valves during this time.
- After zero adjustment, press ▲ to return to the DP display page and observe it for a moment. If the average DP value is within ±5, the zero adjustment is complete.
- If it is not within this range, repeat the process starting from Step 1.
Menu terminology may vary slightly depending on the display model; refer to the instruction manual included with the meter.
After zero adjustment, open the upstream valve first, then the downstream valve to resume normal operation.

Adjusting the Coriolis Flow Meter Coefficient
A customer previously purchased our smallest-diameter Coriolis mass flow meter to measure a fluorinated solvent with a flow rate of only a few hundred milliliters per hour. After using it for some time, he noticed that the meter’s readings did not match the actual consumption, so he consulted us: Is it possible to adjust the mass flow meter’s flow coefficient?
Yes, it can be adjusted. However, we did not instruct him to make the adjustment immediately; instead, we first asked him to confirm three details:
1. Has the zero point been adjusted?
The smaller the flow rate, the greater the impact of the zero point. Many cases of “inaccurate readings” are actually due to a drifting zero point, which can be resolved with a single zero adjustment.
2. Are you comparing based on mass flow or volumetric flow?
A mass flow meter measures mass. When comparing with a graduated cylinder or a metering pump—which measure volume—you must multiply the volume by the medium’s actual density to make a valid comparison. The density of this customer’s solvent is approximately 1.6 times that of water; if calculated based on water, the discrepancy would be significant.
3. Are there air bubbles in the pipe?
The outlet pressure is close to atmospheric pressure, and low-boiling-point liquids tend to form bubbles when the pressure drops. If there are bubbles, the readings will fluctuate and tend to be lower than actual.
Only after ruling out all three of these issues—and if the readings remain consistently high or low—should you consider adjusting the flow coefficient.
New coefficient after correction = Original coefficient * (1 – Original error)
The original error must be calculated as the average of multiple data sets.
For example, if the original coefficient is 14,321 and the calibration error is 0.48% (expressed as a positive or negative value):
Corrected coefficient = 14,321 * (1 – 0.0048) = 14,252 (rounded to the nearest whole number)
For example, if the original coefficient is 14,321 and the calibration error is -0.3% (expressed as a positive or negative value):
Corrected coefficient = 14,321 * (1 + 0.003) = 14,364 (rounded to the nearest whole number)
To make adjustments on the flow transmitter:
- First, write down the original coefficient. If you make a mistake, you can revert the change.
- Press and hold SET, then select “Engineer Menu.”
- Enter the password. The password is in the manual that came with the meter; if you cannot find it, please contact us.
- Press SET to scroll down and locate “Flow Coefficient.”
- Press ▶ to move the cursor, and press ▲ to change the number.
- Press SET to save.
Only modify the flow coefficient. Do not change the density coefficient next to it—that is a separate matter.
FAQ
More Flow Measurement Solutions
Coriolis Mass Flow Meter Technology
Flowmeter Calibration & Recalibration
Complete Guide for Micro Motion Flow Meters
Mass Flow Rate and Volumetric Flow Rate: Key Differences and Conversion
Magnetic Flow Meter Calibration Procedure – User Guide

Zhang Wei, possesses 20 years of experience as an automation instrumentation engineer, specializing in the research, design, installation, commissioning, and maintenance of automation instruments.
Face to various instrument communication protocols (such as Modbus, Profibus, etc.), with solid hardware circuit design and software programming skills (proficient in C language and PLC programming). Has extensive project experience; projects he has led and participated in have all achieved outstanding results, improving product accuracy, reducing costs, and increasing production efficiency.
Possesses excellent communication and coordination skills and a strong team spirit, enabling him to quickly respond to customer needs and provide high-quality automation instrumentation solutions.
Want to see more of our engineering guides in Google Search? Add Sino-Inst as a preferred source.