Home > Temperature Sensors > Type J Thermocouple

Type J Thermocouple

Type J thermocouple

A Type J thermocouple—also known as an iron-constantan thermocouple—consists of a positive leg (JP) made of 99.95% pure iron and a negative leg (JN) made of a copper-nickel alloy known as constantan (55% copper, 45% nickel). While these thermocouples have a theoretical measurement range of -200°C to 1200°C, they are typically used within the 0°C to 600°C range. At 100°C, a Type J thermocouple produces a thermoelectromotive force (EMF) of 5.268 mV.

Type J thermocouples are cost-effective and offer a higher thermoelectromotive force output than Type K thermocouples. They can be used in both oxidizing and reducing atmospheres and exhibit good resistance to corrosion from H₂ and CO. However, because the iron component oxidizes rapidly above 600°C, they are not suitable for use in high-temperature environments.

Sino-Inst supports custom J-type thermocouples in three structures: sheathed, assembled, and explosion-proof. Both single and dual-element versions are available, and wire color coding can be provided in accordance with national standards (GB), ANSI, or IEC specifications.

Technical Specifications

ItemSpecification
Thermocouple typeJ
Conductor materialIron (+) / Copper-Nickel (−)
Temperature range0–600°C; Φ6 and Φ8 sheathed probes up to 750°C short-term
Accuracy, Class I±1.5°C or ±0.4% |t|, whichever is greater
Accuracy, Class II±2.5°C or ±0.75% |t|, whichever is greater
Sheath diameterΦ2, 3, 4, 5, 6, 8 mm (duplex from Φ3)
Protection tube (assembled type)Φ16, Φ20 mm
Sheath material321 stainless steel (1Cr18Ni9Ti)
Insulation resistance≥1000 MΩ·m (500 V DC)
Pressure rating4 MPa for threaded and flanged types; ≤15 MPa for tapered thread type
Ingress protectionIP65 (splash-proof head), IP55 (waterproof head)
StandardIEC 60584
Wire color codeGB standard by default (red +, purple −); ANSI or IEC on request

Class II is our standard supply. If you need Class I, tell us when you order.

Sheath Diameter and Operating Temperature

How hot a sheathed Type J probe can run depends on the sheath diameter. The thinner the sheath, the lower the limit.

Sheath OD (mm)Continuous useShort-term use
2.0400°C500°C
3.0 / 4.0 / 5.0500°C600°C
6.0 / 8.0600°C750°C

For continuous use above 500°C, choose Φ6 or Φ8.

Three Types of J-type Thermocouple Structures

Sheathed Type (WRFK)

Slim, flexible, and responsive. Suitable for applications with limited space or where bent installation is required.

Type J sheathed thermocouple with compression thread and flange fittings

Assembled Type (WRF)

Features a protective tube and allows for disassembly and element replacement. Suitable for long-term, fixed installation on pipelines and tanks.

Type J assembled thermocouple with waterproof and splash-proof heads

Explosion-Proof Type

Flameproof junction box, for use in environments with explosive gases. Confirm explosion-proof rating when ordering.

Explosion-proof Type J thermocouple with threaded mount

Type J Thermocouple Reference Table (mV vs Temperature)

°C°FmV
-40-40-1.961
-20-4-0.995
0320.000
20681.019
501222.585
1002125.269
1503028.010
20039210.779
25048213.555
30057216.327
35066219.090
40075221.848
45084224.610
50093227.393
550102230.216
600111233.102
650120236.071
700129239.132
750138242.281

Reference junction at 0°C (32°F). Values per IEC 60584-1 / ITS-90.

Full Type J reference table, −210 to 760°C in 1°C steps

Read the row for the tens, then the column for the units. Example: 125°C is row 120, column +5.

°C+0+1+2+3+4+5+6+7+8+9
-210-8.095-8.076-8.057-8.037-8.017-7.996-7.976-7.955-7.934-7.912
-200-7.890-7.868-7.846-7.824-7.801-7.778-7.755-7.731-7.707-7.683
-190-7.659-7.634-7.610-7.585-7.559-7.534-7.508-7.482-7.456-7.429
-180-7.403-7.376-7.348-7.321-7.293-7.265-7.237-7.209-7.181-7.152
-170-7.123-7.094-7.064-7.035-7.005-6.975-6.944-6.914-6.883-6.853
-160-6.821-6.790-6.759-6.727-6.695-6.663-6.631-6.598-6.566-6.533
-150-6.500-6.467-6.433-6.400-6.366-6.332-6.298-6.263-6.229-6.194
-140-6.159-6.124-6.089-6.054-6.018-5.982-5.946-5.910-5.874-5.838
-130-5.801-5.764-5.727-5.690-5.653-5.616-5.578-5.541-5.503-5.465
-120-5.426-5.388-5.350-5.311-5.272-5.233-5.194-5.155-5.116-5.076
-110-5.037-4.997-4.957-4.917-4.877-4.836-4.796-4.755-4.714-4.674
-100-4.633-4.591-4.550-4.509-4.467-4.425-4.384-4.342-4.300-4.257
-90-4.215-4.173-4.130-4.088-4.045-4.002-3.959-3.916-3.872-3.829
-80-3.786-3.742-3.698-3.654-3.610-3.566-3.522-3.478-3.434-3.389
-70-3.344-3.300-3.255-3.210-3.165-3.120-3.075-3.029-2.984-2.938
-60-2.893-2.847-2.801-2.755-2.709-2.663-2.617-2.571-2.524-2.478
-50-2.431-2.385-2.338-2.291-2.244-2.197-2.150-2.103-2.055-2.008
-40-1.961-1.913-1.865-1.818-1.770-1.722-1.674-1.626-1.578-1.530
-30-1.482-1.433-1.385-1.336-1.288-1.239-1.190-1.142-1.093-1.044
-20-0.995-0.946-0.896-0.847-0.798-0.749-0.699-0.650-0.600-0.550
-10-0.501-0.451-0.401-0.351-0.301-0.251-0.201-0.151-0.101-0.050
00.0000.0500.1010.1510.2020.2530.3030.3540.4050.456
100.5070.5580.6090.6600.7110.7620.8140.8650.9160.968
201.0191.0711.1221.1741.2261.2771.3291.3811.4331.485
301.5371.5891.6411.6931.7451.7971.8491.9021.9542.006
402.0592.1112.1642.2162.2692.3222.3742.4272.4802.532
502.5852.6382.6912.7442.7972.8502.9032.9563.0093.062
603.1163.1693.2223.2753.3293.3823.4363.4893.5433.596
703.6503.7033.7573.8103.8643.9183.9714.0254.0794.133
804.1874.2404.2944.3484.4024.4564.5104.5644.6184.672
904.7264.7814.8354.8894.9434.9975.0525.1065.1605.215
1005.2695.3235.3785.4325.4875.5415.5955.6505.7055.759
1105.8145.8685.9235.9776.0326.0876.1416.1966.2516.306
1206.3606.4156.4706.5256.5796.6346.6896.7446.7996.854
1306.9096.9647.0197.0747.1297.1847.2397.2947.3497.404
1407.4597.5147.5697.6247.6797.7347.7897.8447.9007.955
1508.0108.0658.1208.1758.2318.2868.3418.3968.4528.507
1608.5628.6188.6738.7288.7838.8398.8948.9499.0059.060
1709.1159.1719.2269.2829.3379.3929.4489.5039.5599.614
1809.6699.7259.7809.8369.8919.94710.00210.05710.11310.168
19010.22410.27910.33510.39010.44610.50110.55710.61210.66810.723
20010.77910.83410.89010.94511.00111.05611.11211.16711.22311.278
21011.33411.38911.44511.50111.55611.61211.66711.72311.77811.834
22011.88911.94512.00012.05612.11112.16712.22212.27812.33412.389
23012.44512.50012.55612.61112.66712.72212.77812.83312.88912.944
24013.00013.05613.11113.16713.22213.27813.33313.38913.44413.500
25013.55513.61113.66613.72213.77713.83313.88813.94413.99914.055
26014.11014.16614.22114.27714.33214.38814.44314.49914.55414.609
27014.66514.72014.77614.83114.88714.94214.99815.05315.10915.164
28015.21915.27515.33015.38615.44115.49615.55215.60715.66315.718
29015.77315.82915.88415.94015.99516.05016.10616.16116.21616.272
30016.32716.38316.43816.49316.54916.60416.65916.71516.77016.825
31016.88116.93616.99117.04617.10217.15717.21217.26817.32317.378
32017.43417.48917.54417.59917.65517.71017.76517.82017.87617.931
33017.98618.04118.09718.15218.20718.26218.31818.37318.42818.483
34018.53818.59418.64918.70418.75918.81418.87018.92518.98019.035
35019.09019.14619.20119.25619.31119.36619.42219.47719.53219.587
36019.64219.69719.75319.80819.86319.91819.97320.02820.08320.139
37020.19420.24920.30420.35920.41420.46920.52520.58020.63520.690
38020.74520.80020.85520.91120.96621.02121.07621.13121.18621.241
39021.29721.35221.40721.46221.51721.57221.62721.68321.73821.793
40021.84821.90321.95822.01422.06922.12422.17922.23422.28922.345
41022.40022.45522.51022.56522.62022.67622.73122.78622.84122.896
42022.95223.00723.06223.11723.17223.22823.28323.33823.39323.449
43023.50423.55923.61423.67023.72523.78023.83523.89123.94624.001
44024.05724.11224.16724.22324.27824.33324.38924.44424.49924.555
45024.61024.66524.72124.77624.83224.88724.94324.99825.05325.109
46025.16425.22025.27525.33125.38625.44225.49725.55325.60825.664
47025.72025.77525.83125.88625.94225.99826.05326.10926.16526.220
48026.27626.33226.38726.44326.49926.55526.61026.66626.72226.778
49026.83426.88926.94527.00127.05727.11327.16927.22527.28127.337
50027.39327.44927.50527.56127.61727.67327.72927.78527.84127.897
51027.95328.01028.06628.12228.17828.23428.29128.34728.40328.460
52028.51628.57228.62928.68528.74128.79828.85428.91128.96729.024
53029.08029.13729.19429.25029.30729.36329.42029.47729.53429.590
54029.64729.70429.76129.81829.87429.93129.98830.04530.10230.159
55030.21630.27330.33030.38730.44430.50230.55930.61630.67330.730
56030.78830.84530.90230.96031.01731.07431.13231.18931.24731.304
57031.36231.41931.47731.53531.59231.65031.70831.76631.82331.881
58031.93931.99732.05532.11332.17132.22932.28732.34532.40332.461
59032.51932.57732.63632.69432.75232.81032.86932.92732.98533.044
60033.10233.16133.21933.27833.33733.39533.45433.51333.57133.630
61033.68933.74833.80733.86633.92533.98434.04334.10234.16134.220
62034.27934.33834.39734.45734.51634.57534.63534.69434.75434.813
63034.87334.93234.99235.05135.11135.17135.23035.29035.35035.410
64035.47035.53035.59035.65035.71035.77035.83035.89035.95036.010
65036.07136.13136.19136.25236.31236.37336.43336.49436.55436.615
66036.67536.73636.79736.85836.91836.97937.04037.10137.16237.223
67037.28437.34537.40637.46737.52837.59037.65137.71237.77337.835
68037.89637.95838.01938.08138.14238.20438.26538.32738.38938.450
69038.51238.57438.63638.69838.76038.82238.88438.94639.00839.070
70039.13239.19439.25639.31839.38139.44339.50539.56839.63039.693
71039.75539.81839.88039.94340.00540.06840.13140.19340.25640.319
72040.38240.44540.50840.57040.63340.69640.75940.82240.88640.949
73041.01241.07541.13841.20141.26541.32841.39141.45541.51841.581
74041.64541.70841.77241.83541.89941.96242.02642.09042.15342.217
75042.28142.34442.40842.47242.53642.59942.66342.72742.79142.855
76042.919

Featured Applications

J-Type is not suitable for: humid or rust-prone environments, or long-term exposure to temperatures above 600°C. Use Type K for these conditions.

Order Guide

When inquiring about Type J thermocouples, please provide the following information to our sales engineers:

  1. Temperature measurement range and medium
  2. Construction: sheathed or assembled
  3. Outer diameter and insertion length
  4. Mounting method and thread specifications
  5. Single or dual element; sheathed or insulated
  6. Wire color standards: GB, ANSI, or IEC
  7. Quantity

FAQ

Type J thermocouple wires come in three standard configurations: IEC, ANSI, and GB. The Chinese national standard uses red for the positive lead and purple for the negative; ANSI uses white for the positive and red for the negative; and IEC uses black for the positive and white for the negative. We can customize all three. Simply specify your preference when placing your order.

No, they cannot be directly interchanged. The materials are different, the thermoelectric potentials differ, and both the instrument’s calibration code and the compensation leads must be changed accordingly.

It is commonly used up to 600°C. Sheathed types with diameters of Φ6 and Φ8 can reach 750°C for short periods. For higher temperatures, use a Type K thermocouple.

You can use the following methods to distinguish them:

  1. Use a magnet to check: the positive terminal (iron) of a Type J thermocouple will be attracted to the magnet, while a Type K thermocouple will not;
  2. Measure the thermoelectric characteristics and compare them with the calibration table;
  3. Observe the wire colors: Type K is typically green or gray, while Type J is white or red;
  4. Check the temperature measurement range: Type K can measure from -200°C to 1,372°C, while Type J ranges from 0°C to 750°C;

There are two steps.
Step 1: Set the multimeter to the resistance range and measure the resistance between the two output terminals of the thermocouple. If there is a reading and the resistance value is not too high, it means the thermocouple is not broken. If the display shows an open circuit, it is broken.
Step 2: Set the multimeter to the DC millivolt range and place the probes in boiling water. At a room temperature of around 25°C, the reading should be approximately 4 mV. If there is no reading or the reading does not change, the probe is faulty.

Technical Support and Cases Share

J Type vs K Type Thermocouple: Difference & How to Choose

For temperatures below 750°C, reducing atmospheres, or when on a tight budget, choose a Type J thermocouple; for higher temperatures, oxidizing atmospheres, or general-purpose temperature measurement applications, choose a Type K thermocouple. Type J and Type K thermocouples have the same factory accuracy, which is ±2.2°C or ±0.75%. Type K can measure up to 1250°C, … Read more

Differences Between Thermocouples: Type S, K, N, J, E, T

Thermocouples have the advantages of simple structure, wide measurement range, and easy processing of output signals. Therefore, thermocouples have become our preferred sensor for temperature measurement. The graduation numbers of thermocouples mainly include S, R, B, N, K, E, J, T, etc. Among them, S, R, and B belong to precious metal thermocouples, and N, … Read more

K Type Thermocouple: Working Principle, Applications, Advantages

A type K thermocouple is a widely used thermocouple in temperature measurement. Type K thermocouples are typically composed of two different metal alloys: a chromium-nickel alloy and an aluminum-nickel alloy. Type K thermocouples are inexpensive, highly accurate, reliable, and have a wide temperature range. Their maximum continuous operating temperature is approximately 1100°C. They provide accurate … Read more

    Request For Quote